Buchi DuMaster D-480 Uživatelský manuál Strana 246

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 356
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 245
Operation Manual D-480 ©BÜCH Labortechnik AG
9 - Repairing the instrument 246
The electrical section is now visible.
Pull out the fuse holder of the appropriate fuse.
Press the fuse out.
Check the fuse:
If the fuse wire...
... is visible: Check whether the wire is broken or not.
... is not visible: Measure fuse continuity.
Make a decision for the further proceeding:
If...
... the fuse wire is broken: Insert a new fuse of the same type.
... no continuity was measured on the fuse: Insert a new fuse of the same type.
... the fuse wire is OK: Insert the old fuse again.
... continuity was measured on the fuse: Insert the old fuse again.
Phase 3: Insert fuses
Proceed as follows:
Make sure that the fuse specifications match the label on the fuse holder.
Insert the fuse into the appropriate fuse holder and tighten the semiconductor relay holding plate
again.
Zobrazit stránku 245
1 2 ... 241 242 243 244 245 246 247 248 249 250 251 ... 355 356

Komentáře k této Příručce

Žádné komentáře