5 – 28
5 Installation
KjelDigester K-446/K-449 Operation Manual
bTasks
WARNING
Risk of acid or toxic fume inhalation.
Serious injuries or death can result.
1. Always run the Scrubber or water jet pump during the digestion process.
2. Operate the Scrubber and KjelDigester only in a fume hood.
3. Close the fume hood door as far as possible.
NOTICE
Risk of wrong power input on the Scrubber.
Wrong input voltage of the Scrubber can damage electronic components.
1. Do only connect a Scrubber with 230V input to the KjelDigester. Output volt-
age of the KjelDigester is 240V only.
1. Connect the Scrubber connection cable on the rear of the KjelDigester.
2. Connect the hosec with the ball joint hose connector to the fume collection tubea.
3. Secure the connection with the catchpot clipb.
4. Shorten the other end of the hosec to keep the hose as short and straight as pos-
sible.
5. Connect the hosec to a Scrubber or water jet pump.
cFollow-on tasks
1. Depending on the configuration continue with following step:
∙ Install the condensate bottle (5–33).