6 – 43Operation Manual KjelDigester K-446/K-449
6 Operation
6.8.4 Set parameters for an automatic digestion
1. Press Method [METH].
2. Select a method number between 1 and 9.
6.8.6 Select a method, (644).
3. Pressc [STEP].
The text [STARTTIME] on the display appears.
4. Set a delayed start time if required or pressc [STEP] to continue.
6.8.7 Set up a delayed start time, (644).
5. Pressa [TEMP] and adjust the temperature by pressingd [+/-].
6. Presse [ENTER] to confirm the new value.
7. Pressb [TIME] and adjust the time by pressingd [+/-].
8. Presse [ENTER] to confirm the new value.
9. Pressc [STEP].
10. Repeat the steps 5 9.
NOTE
If not all of the 4 steps are required, set the time to zero on the remaining steps.
11. Pressa [TIME] and adjust the cooling time to keep the Scrubber running after the
digestion is finished by pressingc [+/-].
NOTE
Set up a time amount big enough that no more fumes are evading from the sample tubes
when the Scrubber stops. BUCHI recommends a starting value of 30minutes.
12. Pressd [ENTER] to confirm the new value.
13. Pressb [STEP].