6 Operation
111
K-375/376/377 Operation Manual, Version B
Peripherals
Select Yes for all present peripherals:
Level sensor for receiver waste container
Level sensor for sample tube waste container
Sampler
Balance
Chiller
Printer
NOTE
Unlike the level sensors for the storage tanks of
H
20, NaOH and H3BO3, the level sensors for the
waste containers have to be activated!
If a chiller is connected to the K-375 it is strictly
necessary to select “Yes” for “Chiller used” in
order to have the cooling water valve permanently
opened.
If a sampler is connected to the instrument, select
the type (1 tray for K-376, 2 trays for the K-377)
and enable or disable the auto cleaning functionality
of the sampler.
If a balance is connected to the instrument, specify
the communication settings for the balance.
the Baud rate
the number of Data bits (7 or 8)
the Parity (none, even or odd)
the number of Stop bits (1, 1.5 or 2)
NOTE
For details about the single parameters, please
refer to the operation manual of the balance.