5 Putting into operation
45
K-375/376/377 Operation Manual, Version B
Connect the dosing unit for the acid to connector i.
Connect the RS232 cable to the sampler (if present) to the corresponding connector b.
Connect any additional perihperals according to the description in figure 5.1.
5.2.2 Connections of the K-376/K-377
(left rear side of the housing)
(right rear side of the housing)
a Power switch K-376/K-377
b Power connection K-376/K-377
c Fuses (2 x 3A)
d RS232 connection to K-375
e Toggle switch (see chapter 8.3)
Fig. 5.2: Electrical connections of the K-376/K-377
NOTICE
Risk of device damage by wrong voltage.
Make sure that the voltage on the socket corresponds to the voltage given on the type plate
of the instrument.
Always connect the device to an earthed socket. External connections and extension cables
must be provided with an earthed conductor lead (3-pole couplings, cable or plug equipment)
as the mains lead has a molded plug, thus avoiding risks due to inadvertent defective wiring.
Make sure that no electric sparks form in the device or its surroundings as they might
damage the instrument.
On the K-376/K-377 sampler
Connect the power cable to the power connection b
Connect the cable to the K-375 device to the RS232 connector d