13
Rotavapor R-250
4 Putting into Operation
Fig. 4.6: Reflux glass assembly
4.7 Installation of the Reflux Glass Assembly
All glassware used must be intact, with no signs of cracks,
spreading impact marks, or other damage. Inspect the glass-
ware visually before installing it.
s &IXTHEREmUXDISTRIBUTIONHEAD햲 in position on the gear
head using a DN 70 EasyClamp connection.
s &ITTHECOOLINGWATERHOSENIPPLES햳 to both coolers 햴.
s )NTRODUCEBOTHCOOLERS햴, together with the cooler hol-
ders 햵, into the pivoting clamps 햶 and connect them to
the distribution head (2 x DN 40 EasyClamp connections).
s !LIGNTHECOOLERS햴 in a vertical position and fix in place
with the pivoting clamps 햶.
s 3CREWTHESHUTOFFTAP햷 into the distribution head 햲
and tighten it firmly.
s )NSERTTHEINLETTAP햸 into the distribution head and tighten
the SVL 30 nut.
s 0LACE THE CONDENSATE COOLER 햹 on the drip tray and
connect it using the EasyClamp connection. Fit the con-
necting pipe 햺 between the condensate cooler and the
distribution head 햲 and secure it (2 x DN 25 EasyClamp
connections).
s 3CREW THE TEMPERATURE SENSOR 햻 into the distribution
head 햲.
s 4IGHTENALL%ASY#LAMPCONNECTIONSEVENLYANDINPARALLEL
s 4URNTHEPOSITIONINGNUTTORAISETHEmASKSUPPORT#ONNECT
the receiving flask to the condensation flask.
햸
햻
햲
햳
햶
햷
햴
햺
햹
햳
햵
햳