Shrnutí obsahu

Strany 1 - Operation Manual

ELS Detector C-650Operation Manual11593555 en

Strany 2

C-650 Operation Manual, Version B2 Safety102.5.2 Warning labels on housing and assembliesThe following warning sticker(s) can be found on the housin

Strany 3 - Table of content

C-650 Operation Manual, Version B3 Technical data113 Technical dataThis chapter introduces the device specifications. It contains the scope of deliv

Strany 4

C-650 Operation Manual, Version B3 Technical data123.1.3 Optional accessoriesTable 3.3: Optional accessoriesProduct Order numberFlow splitter 110 5

Strany 5 - 1 About this manual

C-650 Operation Manual, Version B4 Description of function134 Description of functionThis chapter explains the basic principle of the device, shows

Strany 6 - 2 Safety

C-650 Operation Manual, Version B4 Description of function144.2 Functional principleThere are three discrete steps in the operation of the detector

Strany 7 - 2 Safety

C-650 Operation Manual, Version B4 Description of function151234a Liquid inletb Coloured sealc Ringd Gas inletFig 4.3: Design of the nebulizerFig 4.

Strany 8

C-650 Operation Manual, Version B4 Description of function164.2.2 Evaporation of the solventA heated tube is used to evaporate the solvent. The exi

Strany 9

C-650 Operation Manual, Version B4 Description of function17A plot of log I versus log m provides a linear response. The values of the constants (k

Strany 10 - Operation Manual, Version B

C-650 Operation Manual, Version B5 Putting into operation185 Putting into operationThis chapter describes how the device is installed and gives ins

Strany 11 - 3 Technical data

C-650 Operation Manual, Version B5 Putting into operation195.2 Electrical connectionsNOTEExternal connections and extension lines must be provided

Strany 13 - 4 Description of function

C-650 Operation Manual, Version B5 Putting into operation20The blue cables are contact closure “input” cables that are used to power the unit down (

Strany 14

C-650 Operation Manual, Version B5 Putting into operation21After Parafilm is removed from the detector gas inlet, the tubing should be cut and firmly

Strany 15

C-650 Operation Manual, Version B5 Putting into operation225.4 Exhaust venting and drain requirements WARNINGRisk of toxic fume inhalation. • Alway

Strany 16

C-650 Operation Manual, Version B5 Putting into operation235.5 Installing the nebulizer/glass chamber assemblyParafilmTM is used to cover various op

Strany 17

C-650 Operation Manual, Version B5 Putting into operation24Use the large black nut to position the nebulizer on the glass chamber.Fill the siphon ov

Strany 18 - 5 Putting into operation

C-650 Operation Manual, Version B5 Putting into operation255.6 Installing the flow splitter5.6.1 Connecting the flow splitterThe flow splitter comes

Strany 19

C-650 Operation Manual, Version B5 Putting into operation26flask. This adjustment requires only a small rotation of the knob. Too big rotation could

Strany 20

C-650 Operation Manual, Version B5 Putting into operation27After the detector has completed the initialization procedures, the Status screen (Fig 5.

Strany 21

C-650 Operation Manual, Version B5 Putting into operation28gain can be set via the Temp/Gain screen (Fig 5.16) and the pressure units can be selecte

Strany 22

C-650 Operation Manual, Version B5 Putting into operation29When the detector is powered up or if you change the temperature, the temperature may firs

Strany 23

Table of content1 About this manual 51.1 Abbreviations 52 Safety 62.1 User qualification 62.2 Proper use 62.3 Improper use 62.4 Safety warni

Strany 24

C-650 Operation Manual, Version B5 Putting into operation30Noise Filter/Pressure unit screenFilter: 1 sPress Unit: barFig 5.18: Noise Filter/Pressur

Strany 25

C-650 Operation Manual, Version B5 Putting into operation31 button followed by the button to turn it off.The # hours field indicates the number of

Strany 26

C-650 Operation Manual, Version B5 Putting into operation32The three options provided for external shutdown are summarized in the following table:Mo

Strany 27

C-650 Operation Manual, Version B5 Putting into operation33Serial Number ScreenSerial Number #########Fig 5.23: Serial Num

Strany 28

C-650 Operation Manual, Version B5 Putting into operation34Factory Menu Code ScreenFactory Menu Code ____Authorized persons onlyFig 5.25: Factory Me

Strany 29

C-650 Operation Manual, Version B5 Putting into operation35When the detector is powered up or if you change the temperature, the temperature may firs

Strany 30

C-650 Operation Manual, Version B5 Putting into operation365.9.4 Solvent noise testTo determine the solvent noise: Ensure that the gas is flowing at

Strany 31

C-650 Operation Manual, Version B5 Putting into operation375.9.5 Column noise testNOTEWhen strongly retained compounds are slowly eluted from the c

Strany 32

C-650 Operation Manual, Version B6 Operation386 OperationThis chapter describes the operations that should be performed on a routine basis when you

Strany 33

C-650 Operation Manual, Version B6 Operation396.2.2 ExternalTo auto-zero the detector, a contact closure signal or a TTL signal is used to short cir

Strany 34

5.8 Control panel 265.8.1 Digital display 265.8.2 User interface 275.9 Initial test procedures 345.9.1 Preliminary activities 345.9.2 Elect

Strany 35

C-650 Operation Manual, Version B6 Operation40NOTEIf the mobile phase used is not easily volatile, such as DMSO or DMF, temperature should be increas

Strany 36

C-650 Operation Manual, Version B6 Operation41 Fig 6.27:Chromatogram with variing solvent qualityThe mobile phase should not contain non-vola

Strany 37

C-650 Operation Manual, Version B6 Operation426.4.5 Optimizing the noise filterThe Digital Filter (Fig 5.18) allows maximizing Signal-to-Noise ratio

Strany 38 - 6 Operation

C-650 Operation Manual, Version B7 Maintenance437 MaintenanceThis chapter gives instructions on all maintenance work to be performed in order to kee

Strany 39

C-650 Operation Manual, Version B7 Maintenance44The time required for each step depends on the application, solvents, type and concentration of the s

Strany 40

C-650 Operation Manual, Version B8 Troubleshooting458 Troubleshooting8.1 General troubleshooting informationThe ELSD C-650 Evaporative Light-Scatte

Strany 41

C-650 Operation Manual, Version B8 Troubleshooting468.3 Perform the noise testRepeat the tests described in section 5.9 and compare the observed dat

Strany 42

C-650 Operation Manual, Version B8 Troubleshooting478.5 Nebulizer cleaning and replacement proceduresIf the mist of the nebulizer is not homogeneous

Strany 43 - 7 Maintenance

C-650 Operation Manual, Version B8 Troubleshooting48 • Remove the gas inlet quick fitting and the black plastic nut to avoid damaging the seals with

Strany 44

C-650 Operation Manual, Version B8 Troubleshooting49NOTEIf the nebulizer cleaning procedure does not solve the problem, contact your local distributo

Strany 45 - 8 Troubleshooting

C-650 Operation Manual, Version B1 About this manual51 About this manualThis manual describes the Low Temperature Evaporative Light Scattering Dete

Strany 46

C-650 Operation Manual, Version B8 Troubleshooting508.7.2 Device decontamination • Set the evaporation temperature to 100°C and the gas pressure to

Strany 47

C-650 Operation Manual, Version B9 Shutdown, storage, transport and disposal519 Shutdown, storage, transport and disposalThis chapter instructs how

Strany 48

C-650 Operation Manual, Version B10 Spare parts5210 Spare partsThis chapter lists spare parts, accessories, and options including their ordering in

Strany 49

C-650 Operation Manual, Version B11 Declarations and requirements5311 Declarations and requirements11.1 FCC requirements (for USA and Canada)Engli

Strany 50

C-650 Operation Manual, Version B11 Declarations and requirements5411.2 Declaration of conformity

Strany 52 - 10 Spare parts

11593555 en 1012/Technical data are subject to change without notice/ Quality Systems ISO 9001 The English version is the original language Version Bn

Strany 53

C-650 Operation Manual, Version B2 Safety62 SafetyThis chapter points out the safety concept of the device and contains general rules of behavior an

Strany 54

C-650 Operation Manual, Version B2 Safety72.4 Safety warnings and safety signsDANGER, WARNING, CAUTION and NOTICE are standardized signal words for

Strany 55

C-650 Operation Manual, Version B2 Safety8Symbol MeaningHarmful to life-formsPinch point. Mechanical hazard.Fire hazardHot item, hot surfaceDevice da

Strany 56 - Quality in your hands

C-650 Operation Manual, Version B2 Safety9Additional user informationParagraphs starting with NOTE transport helpful information for working with the

Komentáře k této Příručce