Buchi ELS Detector C-650 Uživatelský manuál Strana 50

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 56
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 49
C-650 Operation Manual, Version B
8 Troubleshooting
50
8.7.2 Device decontamination
Set the evaporation temperature to 100°C and the gas pressure to 2bar (29psi).
Pump the appropriate solvent through the system at the rate of 1mL/min. The solvent will be de-
termined by the nature of the samples that were previously analyzed by the detector. If you do not
know the nature of the sample, ethanol is a good choice. Do not use solvents that can potentially
corrode the device. Maintain the flow and temperature during 3hours at least.
Clean the outside of the detector with a non-abrasive cloth. If necessary, a liquid such as
soapy water or ethanol can be used to remove stains or foreign material.
8.8 Light source exchange
In case the light source needs to be exchanged, please contact the BUCHI Customer Service.
Zobrazit stránku 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 56

Komentáře k této Příručce

Žádné komentáře