5 Putting into operation
35 R-220 SE Operation Manual, Version F
5.3 Sensor setup and adjustments
Sensor setup menu
These instructions are valid for the setup menu of the cooling water sensor and level sensors
connected at the rear connection field.
AERATE
UP
=
LEVEL1
UP
AERATE
=
LEVEL2
UP
AERATE
=
Lift
UP
+
Flow
a
b
c
d
Store selection
Store selection
Store selection and
leave setup menu
Keep the "LIFT UP" button pressed while switching on
the Rotavapor to enter the setup menu.
Press the "UP" button to toggle between 'on' (cooling
water flow sensor activated) and 'off' (sensor deacti-
vated).
Press "AERATE" button to store the selected sensor
setting and automatically proceed with level sensor 1.
Press the "UP" button to toggle between 'on' (level
sensor 1 activated) and 'off' (sensor deactivated).
Press "AERATE" button to store the selected sensor
setting and automatically proceed with level sensor 2.
Press the "UP" button to toggle between 'on' (level
sensor 2 activated) and 'off' (sensor deactivated).
Press "AERATE" button to store the selected sensor
setting and to leave the setup menu.