3 Technical data10 V-700/710 Operation Manual, Version ETable 3-1: Accessories Vacuum Pump V-700/710Product Order numberSet hose connections (not on

Strany 3 - Table of contents

3 Technical data11 V-700/710 Operation Manual, Version EOrder number:0 7x3x xVacuum Controller V-855 100 V–240 V 50/60 HzOrder number:0 7x4x xEasyVac

Strany 4

3 Technical data12 V-700/710 Operation Manual, Version E3.1.3 Two controllersOrder number:0 7x6 2xTwo vacuum controllers V-850, two valve unitOrder

Strany 5 - 1 About this manual

3 Technical data13 V-700/710 Operation Manual, Version E3.1.5 Condensation assemblyOrder number:0 7x x x1Secondary condenser and insulationOrder num

Strany 6 - 2 Safety

3 Technical data14 V-700/710 Operation Manual, Version E3.2 Technical data overviewTable 3-2: Technical dataVacuum Pump V-700 Vacuum Pump V-710Dimen

Strany 7 - 2.5 Product safety

3 Technical data15 V-700/710 Operation Manual, Version E3.4 Solvent tableTable 3-4: Solvent tableSolvent Formula Molar mass in g / molEvaporation en

Strany 8 - 2.6 General safety rules

4 Description of function16 V-700/710 Operation Manual, Version E4 Description of functionThis chapter explains the basic principle of the instrumen

Strany 9 - 3 Technical data

4 Description of function17 V-700/710 Operation Manual, Version E4.1.3 Rear view4456327891 1010a Power supplyb Fan for pump coolingc Pump outletd

Strany 10 - Product Order number

4 Description of function18 V-700/710 Operation Manual, Version E4456327891 1010a Power supplyb Fan for pump coolingc Pump outletd Knu

Strany 11

4 Description of function19 V-700/710 Operation Manual, Version E4.1.4 Gas ballastThe gas ballast serves to aerate the system (membrane and valves)

Strany 13

4 Description of function20 V-700/710 Operation Manual, Version E4.2.2 Rear connections of the vacuum controller123456712811 910a Connection for

Strany 14 - 3.3 Materials used

4 Description of function21 V-700/710 Operation Manual, Version E4.3 Vacuum Module V-801 EasyVacThe Vacuum Pump V-700/710 can be used for the automa

Strany 15 - 3.4 Solvent table

4 Description of function22 V-700/710 Operation Manual, Version E4.3.1 Front view EasyVac123456Fig. 4.8: Front view EasyVaca Detection start of dist

Strany 16 - 4 Description of function

4 Description of function23 V-700/710 Operation Manual, Version EThe LEDs indicate the operating state of the system:c LED ON/OFF• Permanent light:

Strany 17

4 Description of function24 V-700/710 Operation Manual, Version E4.4 Vacuum Module V-802 Lab VacThe Vacuum Pump V-700/710 can be used together with

Strany 18

4 Description of function25 V-700/710 Operation Manual, Version EThe LEDs indicate the operating state of the system:c LED ON/OFF• Permanent light:

Strany 19

4 Description of function26 V-700/710 Operation Manual, Version E4.7 Connections to the secondary condenser123123a Outlet for exhaust gasesb Outlet

Strany 20

5 Putting into operation27 V-700/710 Operation Manual, Version E5 Putting into operationThis chapter describes how the instrument is installed and g

Strany 21

5 Putting into operation28 V-700/710 Operation Manual, Version E5.3 Electrical connectionsATTENTIONMake sure that the voltage on the socket correspo

Strany 22 - Fig. 4.8: Front view EasyVac

5 Putting into operation29 V-700/710 Operation Manual, Version E5.4 Silencer or secondary condenser5.4.1 Secondary condenserTo install the secondar

Strany 23

Table of contents3 V-700/710 Operation Manual, Version ETable of contents1 About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 24 - Fig. 4.11: Front view LabVac

5 Putting into operation30 V-700/710 Operation Manual, Version E5.4.2 SilencerInstall the silencer to the pump outlet or to the secondary condenser o

Strany 25

5 Putting into operation31 V-700/710 Operation Manual, Version E• Introduce the Woulff bottle to the holder.• Screw the GL-14 nut of the Woulff b

Strany 26

5 Putting into operation32 V-700/710 Operation Manual, Version E5.7 Complete system connection with one Rotavapor and one vacuum controller123456789

Strany 27 - 5 Putting into operation

5 Putting into operation33 V-700/710 Operation Manual, Version E5.8 System pump with two controllers on two RotavaporFig. 5.8: Complete system conne

Strany 28 - 5.3 Electrical connections

5 Putting into operation34 V-700/710 Operation Manual, Version E5.9 System pump with two controllers on one pumpOption 1Option 1 is a variant with b

Strany 29 - Fig. 5.3: Secondary condenser

5 Putting into operation35 V-700/710 Operation Manual, Version EThe following installation instruction is valid for both variants, although the contr

Strany 30 - Fig. 5.4: Silencer

5 Putting into operation36 V-700/710 Operation Manual, Version E Commissioning option 1 (parallel)To install the controllers proceed as follows:• 1.

Strany 31

5 Putting into operation37 V-700/710 Operation Manual, Version ECommissioning option 2 (opposite)To install the controllers proceed as follows:• 1.

Strany 32

5 Putting into operation38 V-700/710 Operation Manual, Version ECablingHow to connect the cables:1. Use the two green power supply cables to connect

Strany 33

6 Operation39 V-700/710 Operation Manual, Version E6 OperationThis chapter explains the operating elements and possible operating modes. It gives inst

Strany 34 - Option 2

Table of contents4 V-700/710 Operation Manual, Version E4.6 Vacuum connections to the valve unit or Woulff bottle. . . . . . . . . . . . . . . . . 25

Strany 35

6 Operation40 V-700/710 Operation Manual, Version E6.2 Selecting the distillation conditionsTo achieve optimal distillation conditions, the distillat

Strany 36

6 Operation41 V-700/710 Operation Manual, Version E6.3 Optimizing the distillation conditionsDepending on the solvent being distilled the distillatio

Strany 37

7 Maintenance and repairs42 V-700/710 Operation Manual, Version E7 Maintenance and repairsThis chapter gives instructions on all maintenance work to

Strany 38

7 Maintenance and repairs43 V-700/710 Operation Manual, Version E7.4 Valve head7.4.1 CleaningIf the end vacuum of 10 mbar is not reached anymore and

Strany 39 - 6 Operation

7 Maintenance and repairs44 V-700/710 Operation Manual, Version E• Remove the two rubber parts at the valve in- and outlet.• Loosen the black tub

Strany 40 - Fig. 6.1: 20-40-60 ° rule

7 Maintenance and repairs45 V-700/710 Operation Manual, Version E11• In case the valve has to be replaced, loosen the two screws a by means of the

Strany 41

7 Maintenance and repairs46 V-700/710 Operation Manual, Version E• When all replacement work is done, reas-semble the pump head in the reverse orde

Strany 42 - 7 Maintenance and repairs

8 Troubleshooting47 V-700/710 Operation Manual, Version E8 TroubleshootingThis chapter helps to resume operation after a minor problem has occurred wi

Strany 43 - 7.4 Valve head

9 Shutdown, storage, transport and disposal48 V-700/710 Operation Manual, Version E9 Shutdown, storage, transport and disposalThis chapter instructs

Strany 44

9 Shutdown, storage, transport and disposal49 V-700/710 Operation Manual, Version E9.3 Health and safety clearance formDeclaration concerning safety

Strany 45 - Check-valves V-710

1 About this manual5 V-700/710 Operation Manual, Version E1 About this manualThis manual describes the vacuum pump V-700 and V-710 and provides all

Strany 46 - 7.6 Customer service

10 Spare parts50 V-700/710 Operation Manual, Version E10 Spare parts and accessoriesThis chapter lists spare parts, accessories, and options includi

Strany 47 - 8 Troubleshooting

10 Spare parts51 V-700/710 Operation Manual, Version ETable 10-1: Spare parts (cont.)Product Order numberControl cable between vacuum controller and

Strany 48 - 9.2 Disposal

10 Spare parts52 V-700/710 Operation Manual, Version ETable 10-1: Spare parts (cont.)Product Order numberWoulff bottle complete (with R-210/215 and V

Strany 49

10 Spare parts53 V-700/710 Operation Manual, Version ETable 10-1: Spare parts (cont.)Product Order numberSet of 4 valves, including O-rings and screw

Strany 50 - 10.1 Spare parts

11 Declarations and requirements54 V-700/710 Operation Manual, Version E11 Declarations and requirements11.1 FCC requirements (for USA and Canada)E

Strany 51

11 Declarations and requirements55 V-700/710 Operation Manual, Version E11.2 Declaration of conformity

Strany 54

BÜCHI Labortechnik AGCH-9230 Flawil 1 / SwitzerlandT +41 71 394 63 63 F +41 71 394 65 65 www.buchi.com Quality in your hands

Strany 55

2 Safety6 V-700/710 Operation Manual, Version E2 SafetyThis chapter points out the safety concept of the instrument and contains general rules of beha

Strany 56

2 Safety7 V-700/710 Operation Manual, Version E2.4 Warning notices used in this manualWARNINGGenerally, the triangular warning symbol indicates the p

Strany 57

2 Safety8 V-700/710 Operation Manual, Version E2.6 General safety rulesResponsibility of the operatorThe head of laboratory is responsible for traini

Strany 58 - F +41 71 394 65 65

3 Technical data9 V-700/710 Operation Manual, Version E3 Technical dataThis chapter introduces the reader to the instrument specifications. It contai

Příbuzné modely Vacuum Pump V-700

Komentáře k této Příručce